UNIT CONTROLS
Panel Cover: A cover which conceals the controls for the Pulse Width, Pulse
Rate, and Mode Selector. Press the top portion of the cover and pull down in
order to open the cover.
Intensity: The intensity knobs located on the top of the unit affect the
strengthof thestimulationandalsofunctionasON/OFFcontrols.
Mode: The Mode switch is used to select the type of treatment utilized. The
three modes are Burst (B), Continuous (C), and Modulation (M).
Pulse Width: The Pulse Width knob regulates the pulse width for both
channels.
Pulse Rate: The Pulse Rate knob regulates the number of pulses per second
for both channels.
Mode Functions:
Burst (B) releases individual bursts twice per second, pulse width is
adjustable and the pulse rate is set at 100Hz per second.
Continuous (C) stimulation is delivered continuously at the settings
determined by intensity, rate, and width knobs.
Modulation (M) pulse width decreases the pulse width down to 60% of the
original width setting. This decreased pulse width is maintained for 1.5
seconds before returning to the original width setting, which is maintained
for 3.5 seconds. The cycle is then repeated. The intensity and pulse rate are
adjustable.