Operation
62
Spray arm pressure and monitoring
The machine has a sensor for monitoring spray pressure e.g in order
to detect pressure fluctuations caused by improper loading of the
machine or foam in the water circulation system. Spray pressure
monitoring is activated for the “Cleaning” and “Final rinse” wash
blocks.
The spray pressure monitoring result is documented together with the
process documentation.
Spray arm speed can also be monitored, e.g. for detection of block-
ages caused by improper loading of the machine or foam in the water
circulation system. Spray arm speed monitoring can be activated or
deactivated via the programmable settings.
Miele Service can make further settings for spray pressure and spray
arm monitoring.