Program settings
108
Activating the
conductivity sen-
sor
The conductivity is measured during the final rinse phase (see pro-
gram charts).
...
Final rinse 1 or Final rinse 2
Conductivity Monitoring
Conductivity Monitoring
Yes
No
- Yes
The conductivity is being measured.
- No
The conductivity sensor is deactivated.
Select an option using the and arrow buttons.
Press OK to save the setting.
Changing water
quantity
Increasing the water level is advisable if a large amount of water
clings to items due to the structure of the wash load or if a heavy
build-up of foam might occur due to the type of soiling (e.g. blood)
and the process chemicals used. The additional amount of water re-
quired depends on the type of basket or mobile unit used, the type of
soiling and the load.
If a lightly soiled load is being reprocessed which does not hold much
water, the amount of water can be reset to the factory default amount
to save water and energy.
...
Water volume change
Water volume change
0.0
l
(0.0 - 8.5 l)
The water quantity can be increased in 17 oz. (0.5l) increments, or
set back to the factory default amount. The possible range is shown
in the bottom line. The setting “0l” equates to the factory default set-
ting.
Use the arrow buttons (higher) and (lower) to alter the water
quantity.
Press OK to save the setting.