Program settings
107
Repeating if limit
value is exceeded
If the conductivity for incoming water or draining water exceeds the
limit value, the measurement can be repeated.
If the limit value for incoming water is too high, the water is pumped
out and fresh water is allowed in again. If the conductivity value at the
end of the final rinse phase exceeds the limit value for draining water,
the entire final rinse block is repeated.
If the measured value still exceeds the limit value after the repeat
measurement, the program ends and a fault message is generated.
The message can be shown on the display and also in the cycle re-
port.
...
Conductivity Limit
Water intake and/or Water drainage
Number of repeats
Number of repeats
1
(0 - 1)
The setting value is entered in increments of 1. The possible range is
shown in the bottom line of the display.
Use the arrow buttons (higher) and (lower) to set the number of
repetitions.
Press OK to save the setting.
Calibrating the
conductivity me-
ter
At regular intervals, e.g. in the course of maintenance, the conductiv-
ity meter must be calibrated by Miele Service. Calibration outside the
maintenance cycles may also become necessary.
Conductivity module
requires calibration
If this message appears, contact Miele Service.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Miele PG 8583 and is the answer not in the manual?

Miele PG 8583 Specifications

General IconGeneral
CategoryLaboratory Equipment
ModelPG 8583
TypeLaboratory glassware washer
External dimensions, height835 mm
External dimensions, width600 mm
Usable chamber depth500 mm
Number of baskets2
Max temperature93°C
HeatingElectric
Water connectionCold and hot water
Electrical connection230V, 50Hz
Power supply230V, 50Hz

Related product manuals