Program settings
106
Activating con-
ductivity measur-
ing
Conductivity measuring is activated individually for each program.
Conductivity is measured in the final rinse phase.
...
Conductivity Limit
Conductivity Limit
Water intake
Water drainage
- Water intake
The electrical conductivity of the water after water intake during fi-
nal rinse phase is measured.
- Water drainage
The electrical conductivity of the water at the end of the final rinse
phase is measured.
Select an option using the and arrow buttons.
Press OK to save the setting.
Following selection a maximum limit value for electrical conductivity
must be set.
Setting limit val-
ues
The limit values for Water intake and Water drainage are set individu-
ally.
...
Conductivity Limit
Water intake and/or Water drainage
Set
Water intake
--
µS/cm
(5 - 200 µS/cm)
--
µS/cm
(5 - 200
µS/cm)
Water drainage
The limit values are set in increments of 1. The possible range is
shown in the bottom line of the display.
Use the arrow buttons (higher) and (lower) to set the limit
value.
Press OK to save the limit value.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Miele PG 8583 and is the answer not in the manual?

Miele PG 8583 Specifications

General IconGeneral
CategoryLaboratory Equipment
ModelPG 8583
TypeLaboratory glassware washer
External dimensions, height835 mm
External dimensions, width600 mm
Usable chamber depth500 mm
Number of baskets2
Max temperature93°C
HeatingElectric
Water connectionCold and hot water
Electrical connection230V, 50Hz
Power supply230V, 50Hz

Related product manuals