1 - 1
1 Safety
1.1 Safety Information
• Indicates an imminent hazard that, if not avoided, will result in death or serious injury.
• Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious
injury.
• Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal
injury or product/property damage.
• Provides application tips or other useful information to ensure that you get the most from your
product.
1.1.1 Dangers
• The equipment delivers up to 360 J of electrical energy. Unless properly used by following the
prompts provided by the equipment, this electrical energy may cause serious injury or death. Do not
attempt to operate this equipment unless thoroughly familiar with the operations and functions of
all controls, indicators, connectors, and accessories.
• To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich atmospheres,
flammable anesthetics, or other flammable agents (such as gasoline). Keep the equipment and the
operating environment dry and clean.
• Defibrillation current can cause operator or bystander severe injury or even death. Keep distance
from the patient or metal devices connected to the patient during defibrillation.
1.1.2 Warnings
• Only use parts and accessories specified in this manual. Follow the instructions for use and adhere to
all warnings and cautions.
• This equipment is used for single patient at a time.
• The equipment is not intended to be used within the Magnetic Resonance (MR) environment.
• Do not disassemble the equipment. It contains no operator serviceable components and dangerous
high voltages may be present. Contact authorized service personnel for repair.
• Before connecting the equipment to the external power supply, check that the voltage and
frequency ratings are the same as those indicated on the equipment’s label or in this manual.
• Before each use, the operator must check the equipment condition to ensure that the equipment is
ready for operation.