A - 3
A.2.3 Charger Station Specifications
Battery charge time Charged by the equipment
connected to the external power
supply
Less than 3 hours to 90% and less than 4 hours to
100% with equipment turned off.
Less than 5 hours to 90% and less than 6 hours to
100% with equipment turned on.
Charged by the charger station Less than 3 hours to 90% and less than 4 hours to
100%.
Battery run time Operating
mode
One battery Testing condition
Monitor ≥6.5 h New fully charged battery.
Ambient temperature of 20ºC±5 °C.
The equipment is configured with 3-/5-lead ECG,
manual defibrillation, screen brightness set to the
lowest level without printing.
≥5.5 h New fully charged battery.
Ambient temperature of 20ºC±5 °C.
The equipment is configured with 3-/5-lead ECG,
Resp, SpO2, NIBP measurements set at an interval
of 15 minutes, manual defibrillation, screen
brightness set to the factory default without
printing.
Defibrillation ≥220 discharges
of 360J
New fully charged battery.
Ambient temperature of 20ºC±5 °C.
The equipment is configured with 3-/5-lead ECG,
manual defibrillation, screen brightness set to the
factory default without printing.
3 discharges every minute.
≥300 discharges
of 200J
Pacing ≥4.5 h New fully charged battery.
Ambient temperature of 20ºC±5 °C.
The equipment is configured with 3-/5-lead ECG,
manual defibrillation, external pacing, screen
brightness set to the factory default without
printing.
50  load impedance, pacer rate at 80bpm, pacer
output at 60mA, pacer pulse at 40ms.
Low alarm
presented
At least 20
minutes of
monitoring and
6 discharges of
360J
Ambient temperature of 20ºC±5 °C.
The equipment is configured with 3-/5-lead ECG,
manual defibrillation, screen brightness set to the
factory default without printing.
Battery fuel gauge 5 LEDs indicating the current battery charge level
Shutdown delay At least 20 minutes of monitoring and 6 discharges of 360J
Maximum number of batteries charged Two batteries can be charged at the same time.
AC power input
Input voltage 100 to 240 VAC (±10%)
Input current 1.8 to 0.8A
Frequency 50/60Hz (±3Hz)
DC power input
Input voltage 12.4 to 30.3 VDC
Input current 15.5 to 6.5A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D20 and is the answer not in the manual?

Mindray BeneHeart D20 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D20
CategoryMedical Equipment
LanguageEnglish

Related product manuals