A - 2
A.2 Power Supply Specifications
A.2.1 External Power Supply Specifications
A.2.2 Battery Specifications
Complies with requirements for medical devices of 6.3.4.2, EN1789 (10.1.3, IEC60601-1-12):
Peak acceleration: 1000m/s
2
(102g)
Duration: 6ms
Pulse shape: half-sine
Number of shocks: 3 shocks per direction per axis (18 shocks in total)
Bump
Complies with the requirements for road ambulances of 6.3.4.2, EN1789.
Peak acceleration: 15g
Pulse duration: 6ms
Number of bumps: 1000
Direction: vertical, with the equipment in its normal operating position.
Vibration
Complies with requirements for medical devices of 6.3.4.2, EN1789 (10.1.3, IEC60601-1-12).
10 Hz to100 Hz: 5.0 (m/s
2
)
2
/Hz
100 Hz to 200 Hz: -7 dB/Octave
200 Hz to 2 000 Hz: 1.0 (m/s
2
)
2
/Hz
Duration: 30 minutes per direction per vertical axis (3 axises in total)
Free fall
1 fall on each surface (6 surfaces in total), at the height of 0.75 m
CO
2
module
Item Tempe ra tu re Relative humidity Barometric
Operating condition 5ºC to 40ºC 5% to 95%,
non-condensing
430mmHg to 790mmHg
(57.3kPa to 105.3kPa)
Storage condition -20ºC to 60ºC
AC power input
Input voltage 100 to 240 VAC (-15%, +10%)
Input current 1.8 to 0.8A
Frequency 50/60Hz (±3Hz)
DC power input (with DC/AC inverter)
Input voltage 12 VDC (-15%, +25%)
Output voltage 230 VAC, 50Hz
Output power 150W
Battery type Rechargeable lithium-ion battery
Battery voltage 14.4V
Battery capacity 4500 mAh
Maximum number of
batteries configured
Only one battery can be connected.