S-6
10Before examining a new patient, press Patient to delete the
patient information and data recorded in the memory for the
previous patient. Otherwise, the new data may be confused with the
data of the previous patient.
11. Do not pull out plugs of the system and its accessories without
turning OFF the power. Otherwise it may result in equipment damage
or even electric shock.
12. Do not turn OFF the power supply of the system during printing,
saving, or invoking. Otherwise it may result in abnormality of these
processes.
NOTE: 1. Do not use the system in the vicinity of strong electromagnetic field (such as
the transformer), which may affect the performance of the monitor.
2. Do not use the system in the vicinity of high-frequency radiation source
(such as the cellular phone), which may affect the performance of the
machine or even lead to failure.
3. To avoid damaging the system, do not use the system in following
environment:
(1) Locations exposed to direct sunlight
(2) Locations subject to sudden changes in temperature
(3) Dusty locations
(4) Locations subject to vibration
(5) Locations near heat generators
(6) Locations with high humidity
4. Turn ON the system only after the power has been OFF for more than 10
seconds. If the system is turned ON immediately after being turned OFF, it
may result in malfunction of the system.
5. Turn OFF the auxiliary power switch or stop transmission through the
[Freeze] key before connecting or disconnecting a transducer. If a
transducer is connected or disconnected with an image displayed, the
system and/or the transducer may malfunction.
6. After using the transducer, remove the ultrasound gel on it and place the
transducer on the transducer holder. Otherwise, water in the gel may enter
the acoustic lens, thus adversely affecting the performance and safety of the
transducer.

Table of Contents

Other manuals for Mindray DP-8800plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DP-8800plus and is the answer not in the manual?

Mindray DP-8800plus Specifications

General IconGeneral
BrandMindray
ModelDP-8800plus
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals