UES-40
Product Specification 1-6 ISSUE4
2-2-2 High-frequency-output monitor
Detects whether the output value is correct for the output setting value. If problems are
detected, the following operations are performed.
• Error code is displayed on output indicator.
• Audible warning is issued.
• High-frequency output is stopped.
2-2-3 Output timing monitor (for monopolar or saline systems only)
Detects the high-frequency output timing. If output continues for more than 60 seconds, the
following operations are performed.
• Error code is displayed on output indicator.
• Audible warning is issued.
• High-frequency output is stopped.
2-2-4 Self-check monitor
Verifies the correct operation of all safety functions. If problems are detected, the following
operations are performed.
• Error code is displayed on output indicator.
• Audible warning is issued.
• Various settings are disabled.
• High-frequency output is stopped.
2-2-5 Control switch monitor
Detects short-circuits of each switch (foot switch and handpiece). If problems are detected,
the following operations are performed.
• Error code is displayed on output indicator.
• Audible warning is issued.
• Various settings are disabled.
• High-frequency output is stopped.
2-3 Additional features
2-3-1 Preset function
Use of the preset function enables the most-recently used output setting, as well as up to
five configurations of preset user settings, to be stored for future use.
2-3-2 CQM function
When split type patient plates are attached, the CQM indicator lights up when there is a
good connection with the patient.