UES-40
Disassembly and Reassembly Procedure 3-4 ISSUE4
2-2 Disassembly procedure
2-2-1 UES-40 main-unit disassembly
(1) Remove the 15 SCREWS holding the TOP
COVER (HCBK3x6SA: 14 SCREWS;
CBK3x6SA: 1 SCREW) and remove the TOP
COVER.
Phillips screwdriver (No. 2)
Note: During reassembly, make sure no
CABLES are caught under the
COVER.
(2) Detach the E45F CABLE from J5 of
UPHF40CR00U (control board).
(3) Detach the E23F CABLE from J5 of
UPHF40SS00U (monitor board).
TOP COVER
CBK3x6SA
1 SCREW
HCBK3x6SA
12 SCREWS (6 on the opposite side)
HCBK3x6SA
2 SCREWS
E45F CABLE
J5
UPHF40CR00U
(control board)
UPHF40SS00U
(monitor board)
J5
E23F CABLE