Detailed information about your system
9
Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouchSelect Plus Flex™
Meter.
Target
Glucose
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coecient
of Variation
(%)
2.22 2.57 0.09 3.58
3.61 3.91 0.11 2.69
6.67 6.86 0.15 2.15
11.10 10.73 0.20 1.83
19.43 19.43 0.41 2.12
Results show that the greatest variability observed
between test strips when tested with blood is 0.25mmol/L
SD or less at glucose levels less than 5.55mmol/L, or 4.5%
CV or less at glucose levels at 5.55mmol/L or above.