PicoScope 2000 Series Programmer's Guide 7
Copyright © 2006–2016 Pico Technology Ltd. All rights reserved. ps2000pg.en r11
3 Sampling modes
PicoScope 2000 Series PC Oscilloscopes can run in various sampling modes.
Block mode. At the highest sampling rates, the oscilloscope collects data much
faster than a PC can read it. In this case, the oscilloscope stores a block of data in
an internal memory buffer, delaying transfer to the PC until the required number of
data points have been sampled.
ETS mode. In this mode, it is possible to increase the effective sampling rate of the
scope when capturing repetitive signals. It is a modified form of block mode.
Streaming modes. At all but the highest sampling rates, these modes allow
accurately timed data to be transferred back to the PC without gaps. The computer
instructs the oscilloscope to start collecting data. The oscilloscope then transfers
data back to the PC without storing it in its own memory, so the size of the data set
is limited only by the size of the PC's memory. Sampling intervals from less than
one microsecond (depending on model) to 60 seconds are possible. There are two
streaming modes:
· Compatible streaming mode
· Fast streaming mode
3.1
Block mode
In block mode, the computer prompts the oscilloscope to collect a block of data in its
internal memory. When the oscilloscope has collected the whole block, it signals that it
is ready and then transfers the whole block to the computer's memory through the
USB port.
The maximum number of values depends upon the size of the oscilloscope's memory.
A PicoScope 2000 Series oscilloscope can sample at a number of different rates that
correspond to the maximum sampling rate divided by 1, 2, 4, 8 and so on.
There is a separate memory buffer for each channel. When a channel is unused, its
memory can be borrowed by the enabled channels. This feature is handled
transparently by the driver.
The driver normally performs a number of setup operations before collecting each
block of data. This can take up to 50 milliseconds. If it is necessary to collect data with
the minimum time interval between blocks, avoid calling setup functions between calls
to ps2000_run_block, ps2000_ready, ps2000_stop and ps2000_get_values.
See Using block mode for programming details.
3.1.1
Using block mode
This is the general procedure for reading and displaying data in block mode:
1. Open the oscilloscope using ps2000_open_unit.
2. Select channel ranges and AC/DC coupling using ps2000_set_channel.
3. Using ps2000_set_trigger, set the trigger if required.
4. Using ps2000_get_timebase, select timebases until you locate the required
time interval per sample.
5. Start the oscilloscope running using ps2000_run_block.
6. Poll the driver to find out if the oscilloscope has finished collecting data, using
ps2000_ready.
7. Transfer the block of data from the oscilloscope using ps2000_get_values or
ps2000_get_times_and_values.

Other manuals for pico Technology PicoScope 2200A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the pico Technology PicoScope 2200A Series and is the answer not in the manual?

pico Technology PicoScope 2200A Series Specifications

General IconGeneral
Brandpico Technology
ModelPicoScope 2200A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals