PicoScope 2000 Series Programmer's Guide 37
Copyright © 2006–2016 Pico Technology Ltd. All rights reserved. ps2000pg.en r11
5.20.1
PS2000_TRIGGER_CONDITIONS structure
A structure of this type is passed to ps2000SetAdvTriggerChannelConditions in
the conditions argument to specify the trigger conditions, and is defined as follows:
typedef struct tPS2000TriggerConditions
{
PS2000_TRIGGER_STATE channelA;
PS2000_TRIGGER_STATE channelB;
PS2000_TRIGGER_STATE channelC;
PS2000_TRIGGER_STATE channelD;
PS2000_TRIGGER_STATE external;
PS2000_TRIGGER_STATE pulseWidthQualifier;
} PS2000_TRIGGER_CONDITIONS;
Applicability
See ps2000SetAdvTriggerChannelConditions
Members
channelA, channelB: the type of condition that should be applied
to each channel. Use these constants:
CONDITION_DONT_CARE (0)
CONDITION_TRUE (1)
CONDITION_FALSE (2)
channelC, channelD: not used
external: not used
pulseWidthQualifier: the type of condition to apply to the pulse
width qualifier. Choose from the same list of constants given under
channelA, channelB.
Remarks
The channels that are set to CONDITION_TRUE or CONDITION_FALSE must all meet
their conditions simultaneously to produce a trigger. Channels set to
CONDITION_DONT_CARE are ignored.
The PicoScope 2202 oscilloscope can use only a single input channel (either channel A
or channel B) for the trigger source. Therefore you may define CONDITION_TRUE or
CONDITION_FALSE for only one of these channels at a time. You can, optionally, set
up the pulse width qualifier in combination with one of the input channels.
The PicoScope 2204, 2204A, 2205 and 2205A models can all trigger from both channel
A and channel B, and therefore all support logic triggering.

Other manuals for pico Technology PicoScope 2200A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the pico Technology PicoScope 2200A Series and is the answer not in the manual?

pico Technology PicoScope 2200A Series Specifications

General IconGeneral
Brandpico Technology
ModelPicoScope 2200A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals