© 2012 by Proceq SA 27
9. Technical specifications
9.1. General
Dimensions 170 x 200 x 45 mm (6.7 x 7.9 x 1.8")
Weight 780 g (27.5 ounces) plus battery pack 120 g (4.2 ounces)
Housing shock resistant ABS plastic
Display large QVGA LCD with adjustable contrast and backlight
Integrated memory flash 32 MB (non-volatile), RAM 64 MB (volatile)
Internal data storage 100’000 - 1’000’000 values (depending on data type)
Interface Ethernet, USB 1.1, RS 232
Battery type rechargeable Li ion or 3 standard size “C” cells
Battery nominal voltage 3.7 V
Battery nominal capacity 4.3 Ah
Battery max. charge voltage 4.2 V
Input voltage 9 to 16 V DC
Max. input current @ 12 V 1 A
Max. input current @ 9.4 V 1.3 A
Battery operation period typically 10 hours
Operating temperature 0 to + 50 °C (32 to 122 °F)
Humidity non-condensing, 90 % max
9.2. Equotip 3 impact devices
Measuring range 1-999 HL
Measuring accuracy ± 4 HL (0.5 % at 800 HL)
Resolution 1 HL; 1 HV; 1 HB; 0.1 HRA; 0.1 HRB; 0.1 HRC; 0.1 HS; 1 MPa
(N/mm
2
)
Impact direction automatic compensation (excl. DL probe)
Impactenergy •11.5NmmforD,DC,E,Sprobes
•11.1NmmforDLprobe
•3.0NmmforCprobe
•90.0NmmforGprobe
Massofimpactbody •5.45g(0.2ounces)forD,DC,E,Sprobes
•7.25g(0.26ounces)forDLprobe
•3.1g(0.11ounces)forCprobe
•20.0g(0.71ounces)forGprobe
Ballindenter •tungstencarbide,3.0mm(0.12")diameterforC,D,DCprobes
•tungstencarbide,2.78mm(0.11")diameterforDLprobe
•tungstencarbide,5.0mm(0.2")diameterforGprobe
•ceramics,3.0 mm (0.12") diameter for S probe
•polycrystallinediamond,3.0 mm (0.12") diameter for E probe