Proton Products InteliSENS DG2030 Diameter Gauge Instruction Manual (Issue 1a)
The data stream can be terminated by sending an escape character to the gauge (ASCII code
decimal 27 or the ‘Esc’ key or the ‘Ctrl’’[‘ key combination on a PC keyboard).
Reading a continuous stream of a block of output parameters
Use the following request format to read a contiguous block of output parameter values as a
continuous stream:
Request
format
#[Output parameter number][Space][Number of parameters]<CR><LF>
Response
format
[Output parameter value]<CR><LF>
Only consecutive parameter values may be requested in a block parameter request.
The update rate for streaming output parameter values depends on both the data transfer rate and
the data length. For a data transfer rate of 1200 bit / s and a data length of seven 8-bit characters per
parameter value, approximately 10 individual parameter values can be transferred per second.
The data stream can be terminated by sending an escape character to the gauge (ASCII code
decimal 27 or the ‘Esc’ key or the ‘Ctrl’’[‘ key combination on a PC keyboard).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Proton InteliSENS DG2030 and is the answer not in the manual?

Proton InteliSENS DG2030 Specifications

General IconGeneral
BrandProton
ModelInteliSENS DG2030
CategoryLaundry Accessories
LanguageEnglish

Related product manuals