RIGOL
X DS6000 Performance Verification Guide
Document Overview
This manual guides users to correctly test the performance of RIGOL DS6000 series digital
oscilloscope.
Main topics in this manual:
Chapter 1 Overview
This chapter introduces the preparations before performing the performance verification tests and
the notices.
Chapter 2 Performance Verification Test
This chapter introduces the limit, test devices required as well as test method and procedures of each
performance specification.
Appendix Test Record Form
The appendix provides a test record form for users to record the test results and judge whether each
performance specification can meet the requirement.
Format Conventions in this Manual:
Front Panel Key: denoted by “Text Box + Button Name (Bold)”, for example, UTIL.
Menu Softkey: denoted by “Character Shading + Menu Word (Bold)”, for example, Self-Cal.
Operation Step: denoted by an arrow “”, for example, UTIL Self-Cal.
Content Conventions in this Manual:
DS6000 series digital oscilloscope includes the following models. In this manual, DS6104 is taken as
an example to illustrate the performance verification test methods. Unless otherwise noted, the
introductions in this manual are applicable to other models.
Model DS6104 DS6102 DS6064 DS6062
Analog Bandwidth 1 GHz 1 GHz 600 MHz 600 MHz
Number of Channels 4 2 4 2
Max Real-time Sample Rate 5 GSa/s
Standard Memory Depth 140 Mpts
Waveform Capture Rate Up to 180 000 wfs/s