RIGOL Chapter 2 Performance Verification Test
2-20 DS6000 Performance Verification Guide
Zero Point Offset Test
Zero point offset is defined as the offset of the crossing point of the waveform and trigger level line
relative to the trigger position as shown in the figure below.
Specification
Zero Point Offset
Specification ±0.5 div × Minimum Time Base Scale
[1]
Note
[1]
: For different models of oscilloscopes under test, the minimum time base scales are different. For DS610X,
the minimum time base scale is 500 ps/div; for DS606X, the minimum time base scale is 1 ns/div.
Test Connection Diagram
Figure 2-6 Zero Point Offset Test Connection Diagram
Test Procedures
1. Connect the active signal terminal of Fluke 9500B to CH1 of the oscilloscope, as shown in the
figure above.
Fluke 9500B
DS6000