Rigol DS6104 Performance Verification Guide

Rigol DS6104
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
RIGOL Chapter 2 Performance Verification Test
2-4 DS6000 Performance Verification Guide
Test Record Form
CH1-CH4 (1MΩ Input Impedance)
Channel Vertical Scale Test Result Limit Pass/Fail
CH1
200 mV/div
0.99 to 1.01
500 mV/div
CH2
200 mV/div
500 mV/div
CH3
200 mV/div
500 mV/div
CH4
200 mV/div
500 mV/div
CH1-CH4 (50 Ω Input Impedance)
Channel Vertical Scale Test Result Limit Pass/Fail
CH1
200 mV/div
49.25 Ω to 50.75 Ω
500 mV/div
CH2
200 mV/div
500 mV/div
CH3
200 mV/div
500 mV/div
CH4
200 mV/div
500 mV/div
External Trigger Channel
Channel Input Impedance Test Result Limit Pass/Fail
EXT
TRIG
1 MΩ 0.99 MΩ to 1.01
50 Ω 49.25 Ω to 50.75 Ω