EasyManuals Logo
Home>Rigol>Test Equipment>DS6104

Rigol DS6104 Performance Verification Guide

Rigol DS6104
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
Chapter 1 Overview RIGOL
DS6000 Performance Verification Guide
1-1
Chapter 1 Overview
Test Preparations
The following preparations should be done before the test.
1. Self-test: perform self-test to make sure that the oscilloscope can work normally;
2. Warm-up: warm the oscilloscope up for at least 30 minutes;
3. Self-calibration: calibrate the oscilloscope.
Self-test
When the oscilloscope is in power-on state, press the power key
at the lower left corner of the
front panel to start the oscilloscope. During the start-up, the instrument performs a series of self-test
items and users can hear the sound of relay switching. The welcome screen is displayed after the
self-test is finished. Users can view the self-test results by pressing UTIL System
SelfTestInfo.
If the self-test fails, make sure that the problems are found and resolved and do not perform
calibration and performance tests until the instrument passes the self-test.
Self-calibration
Make sure that the oscilloscope has been warmed up or running for more than 30 minutes before
performing self-calibration. Then, follow the steps to calibrate the oscilloscope.
1. Connect the [Trig Out/Calibration] connector at the rear panel of the oscilloscope to the four
analog input channels and the external trigger input channel at the front panel respectively using
a one-to-five BNC (M) cable.
2. Press UTIL Self-Cal at the front panel to enter the self-calibration interface.
3. Press Start and the oscilloscope will start to execute the self-calibration.
4. It takes about 20 minutes (15 minutes for dual-channel models) to finish the calibration. Restart
the oscilloscope when “Calibration finished, please restart the oscilloscope!” is displayed.
5. Press ACQ Acquisition to select “Average”; then, press Averages and use to set the
number of averages to 16.
6. Disconnect the input signals of all the channels. Set the vertical scale of each channel to 2 mV/div
and view the offset of the waveform of each channel. If the offset is greater than 1 div, you should
perform self-calibration again.

Other manuals for Rigol DS6104

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS6104 and is the answer not in the manual?

Rigol DS6104 Specifications

General IconGeneral
BrandRigol
ModelDS6104
CategoryTest Equipment
LanguageEnglish

Related product manuals