RIGOL Appendix Test Record Form
DS6000 Performance Verification Guide
DC Gain Accuracy Test
50 Ω Input Impedance
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vavg1 Vavg2 Calculation Result
[1]
CH1
2 mV/div
≤ 2%
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
CH2
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
CH3
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
CH4
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
Note
[1]
: The calculation formula is |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/Full Scale × 100%; wherein, Vout1 and Vout2
are 3 × the current vertical scale and -3 × the current vertical scale respectively.