RIGOL Chapter 2 Command System
2-140 MSO1000Z/DS1000Z Programming Guide
[:SOURce[<n>]]:APPLy:NOISe
[:SOURce[<n>]]:APPLy:PULSe
[:SOURce[<n>]]:APPLy:RAMP
[:SOURce[<n>]]:APPLy:SINusoid
[:SOURce[<n>]]:APPLy:SQUare
[:SOURce[<n>]]:APPLy:USER
Syntax
[:SOURce[<n>]]:APPLy:NOISe [<amp>[,<offset>]]
[:SOURce[<n>]]:APPLy:PULSe [<freq>[,<amp>[,<offset>[,<phase>]]]]
[:SOURce[<n>]]:APPLy:RAMP [<freq>[,<amp>[,<offset>[,<phase>]]]]
[:SOURce[<n>]]:APPLy:SINusoid [<freq>[,<amp>[,<offset>[,<phase>]]]]
[:SOURce[<n>]]:APPLy:SQUare [<freq>[,<amp>[,<offset>[,<phase>]]]]
[:SOURce[<n>]]:APPLy:USER [<freq>[,<amp>[,<offset>[,<phase>]]]]
Description
Configure the specified source channel to output a signal with the specified waveform and
parameters.
Parameter
Name
Type
Range
Default
[<n>]
Discrete
{1|2}
If omitted, the
operation is
performed on
source 1 by default
<freq>
Real
Sine: 0.1Hz to 25MHz
Square: 0.1Hz to 15MHz
Pulse: 0.1Hz to 1MHz
Ramp: 0.1Hz to 100kHz
Arbitrary waveform: 0.1Hz to 10MHz
1kHz
<amp>
Real
Related to the output impedance
HighZ: 20mVpp to 5Vpp
50Ω: 10mVpp to 2.5Vpp
5Vpp
<offset>
Real
Related to the output impedance and
amplitude
HighZ: (-2.5V + current amplitude/2) to
(2.5V - current amplitude/2)
50Ω: (-1.25V + current amplitude/2) to
(1.25V - current amplitude/2)
0μV
DC
<phase>
Real
0° to 360°
0°
Explanation
 These commands are used to select the waveform shape.
NOISe: select Noise
PULSe: select Pulse
RAMP: select Ramp
SINusoid: select Sine
SQUare: select Square
USER: select Arbitrary waveform
 <freq>: set the frequency of the specified waveform (Noise does not have this
parameter). The default unit is Hz;
<amp>: set the amplitude of the specified waveform. The default unit is Vpp;
<offset>: set the DC offset of the specified waveform. The default unit is V
DC
;
<phase>: set the start phase of the specified waveform (Noise does not have this
parameter). The default unit is degree (°).

Table of Contents

Other manuals for Rigol MSO1104Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO1104Z and is the answer not in the manual?

Rigol MSO1104Z Specifications

General IconGeneral
BrandRigol
ModelMSO1104Z
CategoryTest Equipment
LanguageEnglish

Related product manuals