RIGOL Chapter 2 Command System
2-12 MSO1000Z/DS1000Z Programming Guide
:CHANnel<n>:RANGe
Syntax
:CHANnel<n>:RANGe <range>
:CHANnel<n>:RANGe?
Description
Set or query the vertical range of the specified channel. The default unit is V.
Parameter
Name
Type
Range
Default
<n>
Discrete
{1|2|3|4}
--
<range>
Real
Related to the probe ratio
When the probe ratio is 1X: 8mV to 80V
When the probe ratio is 10X: 80mV to 800V
8V (the probe
ratio is 10X)
Explanation
This command indirectly modifies the vertical scale of the specified channel (Vertical Scale
= Vertical Range/8). The vertical scale can be set by the :CHANnel<n>:SCALe command.
Return
Format
The query returns the vertical range in scientific notation.
Example
:CHANnel1:RANGe 8 /*Set the vertical range of CH1 to 8V*/
:CHANnel1:RANGe? /*The query returns 8.000000e+00*/