Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-167
:TRIGger:PULSe:LEVel
Syntax
:TRIGger:PULSe:LEVel <level>
:TRIGger:PULSe:LEVel?
Description
Set or query the trigger level in pulse width trigger. The unit is the same as the current
amplitude unit.
Parameter
Name
Type
Range
Default
<level>
Real
(-5 x VerticalScale - OFFSet) to
(5 x VerticalScale - OFFSet)
0
Explanation
 For VerticalScale, refer to the :CHANnel<n>:SCALe command. For OFFSet, refer to
the :CHANnel<n>:OFFSet command.
 This command is only available when the signal source is an analog channel.
Return
Format
The query returns the trigger level in scientific notation.
Example
:TRIGger:PULSe:LEVel 0.16 /*Set the trigger level to 160mV*/
:TRIGger:PULSe:LEVel? /*The query returns 1.600000e-01*/