cobas c 111 Installation Manual
04906900001
INSTALLATION-MANUAL cobas c 111
Version 3.0 Page 34 of 43
7. Running the analyzer and everyday tasks
Refer to the cobas c 111 Operator’s Manual to see what tasks you must do each day
before running tests and for information on day-to-day running of the analyzer.
8. Shutting down
After completion of work the instrument should be allowed to enter the “Standby”
status and the waste should then be emptied.
You may use the “End of shift” wizard, guiding you through various procedures to
prepare the instrument to be switched off for a longer time, e.g. overnight, weekend, etc.
The instrument can then be switched off by the On/Off switch at the top left side.
After switching off please ensure that the instrument is cleaned according to the
instructions in the Operator Manual to avoid the build up of contamination.
Another way to shutdown the instrument is to press the STOP button, then on the
touch screen press “shutdown”. The instrument will shut down, after that switch off the
instrument by the Power Switch
9. Preparing for transportation
9.1 Drain the instrument with ISE of water and cleaner.
1. Perform the Diagnostic function “Drain ISE Unit”, follow its instructions
2. Store the valve caps and clamps and the electrodes in separate bags at the place of
the calibrator and reference bottles.
3. Now follow all the steps from procedure 9.2 below.
Always use the approved packaging for transportation of the cobas c 111 analyzer.
Before transporting or disposing of an instrument, carry out a Shutting down procedure
and a full decontamination according to local regulations.
9.2 Drain the instrument without ISE of water and cleaner.
1. Remove the nozzle out of the external cleaner bottle and place on a paper tissue.
Close the cleaner bottle and store it at a safe place.
2. Empty the water container. Leave it in place.
3. Perform the diagnostic action “Fill and empty fluid system”.
4. Empty the waste container.
5. Now the fluid tubing can be disconnected.
6. Remove all cuvette segments from the rotor.
7. Remove all reagents from disk, by using the System SW.
8. Remove the reagent disk from the instrument.
Always use the approved packaging for transportation of the cobas c 111 analyzer.
Before transporting or disposing of an instrument, carry out a Shutting down procedure
and a full decontamination according to local regulations.