HRX-OM-O021
Chapter 5 Display and setting of various functions
HRS Series 5.7 Signal for completion of preparation (TEMP READY)
5-19
5.7 Signal for completion of preparation (TEMP
READY)
5.7.1 Signal for completion of preparation (TEMP READY)
This function sets the band width for the circulating fluid set temperature
(upper/lower temperature range) in order to notify the customer by
communication that the circulating fluid temperature has reached the band
range (upper/lower temperature range).The default setting of this function is
“OFF”.
【Tips】
This function is available when contact input/output and serial communication
is used. Refer to the Communications Operation Manual for more details.
An example is shown below.
Circulating fluid set temperature
:20
o
C
READY band width (upper/lower
temperature range)
:±2
o
C
READY time
:60 sec.
Preparation is completed 60 seconds after the circulating fluid temperature
reaches 18
o
C to 22
o
C.
Circulating set fluid 20℃
READY TEMP 22℃
READY TEMP 18℃
READY band width
(upper/lower temperature range) 2℃
READY band width
(upper/lower temperature range) 2℃
Circulating set fluid
READY time
60sec
Completion
of preparation
READY time
lower than 60sec
Contact output:State generation output
Serial:ON output
Contact output:State not generation output
Serial :OFF output
Completion of preparation
(TEMP READY)