Operation
The NAME area will change to MEAS: °FREQ with FREQ underlined by
the cursor.
To select among the various measurements, turn the A control until the
desired measurement is displayed.
The TIME measurement is unique in that you may define unique
measurements by adding a suffix to the °TIME.... label. To change the suffix,
turn the & REF OR DLY POS control untii the cursor underlines the first space
with a " ." in it. Then rotate the A control to obtain the character you want.
This allows you to define multiple TIME measurements which have different
configurations. (For information on configuring the TIME measurement, see
"Time Interval Measurement" in this Section.)
Pressing the MEASURE button toggles back to the NAME: format which
allows saving instrument setups. Subsequent presses of the MEASURE button
toggle between the measurement saving mode and the setup saving mode.
4.
Press one of the setup number buttons. In the upper left corner of the
CRT, the readout will show the number of the button you pushed and
either the new name you defined or the name of the setup previously
associated with that setup number.
Save a
Setup (Extended
Save
setups
9-30)
You can also save more than 8 setups by pressing STEP/AUTO (after
pressing SAVE to enter Save mode), and turning A to the desired extended
setup location. To name the selected setup, follow Step 3 of Save a Setup
(Direct Save setups 1-8), pressing STEP/AUTO when finished. If the name for
the setup is to remain the same, then just press STEP/AUTO.
Save a
Sequence
See Appendix B for sequence programming instructions.
Recall
a
Setup (Direct Recall setups
1-8)
1.
Press RECALL. The readout will indicate the Direct Recall mode by
showing the names of the four setups numbered
1
through 4 in the top
row and the names of the four setups numbered 5 through 8 in the
second row. For example:
Top Row -- SKEW ADJ.PLLTP-2467 CLOCK
Second Row -- ACE KING QUEEN JACK
2465B/2455B/2445B Operators 2-37