EasyManuals Logo
Home>Tektronix>Test Equipment>2465B

Tektronix 2465B Operator's Manual

Tektronix 2465B
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #74 background imageLoading...
Page #74 background image
Applications
Period
and
Frequency
Using Parametric Measurements
To measure Period and Frequency using the Parametric
Measurement feature:
1.
Push the MEASURE button.
2.
Select FREQUENCY from the menu.
The oscilloscope automatically scales the amplitude and timing for the
best measurement. Parametric measurements can only be done on repetitive
signals. Signals with multiple valid trigger points will give erroneous results if
measured with parametric measurements.
Using
the
Counter/Timer/Trigger
(CTT)
1.
Push the MEASURE button.
2.
Select COUNTER from the menu.
3. Select FREQuency from the menu.
Without Parametric Measurements
or CTT
1.
Set SEC/DIV and, if necessary, X10 MAG to spread one cycle over as
wide a span as possible.
2.
Turn on At or
1
/At and align the cursors with identical, fast-slewing
points on consecutive cycles, such as zero-crossings.
3-8
2465B/2455B/2445B Operators

Table of Contents

Other manuals for Tektronix 2465B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2465B and is the answer not in the manual?

Tektronix 2465B Specifications

General IconGeneral
BrandTektronix
Model2465B
CategoryTest Equipment
LanguageEnglish

Related product manuals