EasyManuals Logo
Home>Tektronix>Test Equipment>2465B

Tektronix 2465B Operator's Manual

Tektronix 2465B
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
Applications
Measuring Millivolt Signals
With the standard, 10X-attenuation probes, deflection factors range down to
20 mV/division. To increase the vertical sensitivity by a factor of ten, either use a
1X probe or cascade CH 2 with CH 1. To obtain 200 juV/division, use a 1X probe
or coaxial cable to connect the signal to CH 2 and cascade CH 2 with CH 1. To
cascade CH 2 with CH 1:
1.
Connect the CH 2 output on the rear panel, through a 50-12 cable, to the
CH 1 input on the front panel.
2.
Set the CH 1 input at 1 Mfi DC or AC.
3. Set 20 MHz BW LIMIT on. This will reduce the trace "thickening" caused
by wide-band noise and avoid oscillation of the vertical system. If you
trigger from CH 1 source, you probably need to use HF REJ coupling.
4.
Set CH 2 VOLTS/DIV at 2 mV (20 mV with 10X probe) and set CH 1 at
5 mV or 2 mV/division.
5. Note that the CH 1 scale factor and delta-volts readings agree with the
signal at the CH 2 probe-tip with a 10X probe. With a 1X probe, the scale
factor and delta-volts readings should be divided by ten.
2465B/2455B/2445B Operators
3-17

Table of Contents

Other manuals for Tektronix 2465B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2465B and is the answer not in the manual?

Tektronix 2465B Specifications

General IconGeneral
BrandTektronix
Model2465B
CategoryTest Equipment
LanguageEnglish

Related product manuals