EasyManuals Logo
Home>Tektronix>Measuring Instruments>47E5A

Tektronix 47E5A User Manual

Tektronix 47E5A
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #92 background imageLoading...
Page #92 background image
Characteristics
Sweep Rate (cant)
8 Sweep
Accuracy
+20° C
to
+30° C
-15°
C
to
+55° C
Two-Division Linearity Check
Alternate Sweep Trace Separation
Variable Range (A Only)
82
Table 7 (cont)
Performance Requirements Supplemental Information
HORIZONTAL DEFLECTION SYSTEM (cont)
50
ms
per division
to
0.02
Jl.S
per
division
in
20
steps in a 1-2-5
sequence.
X1
0
MAG
extends
maximum sweep rate
to
2
ns
per
division.
Unmagnified
Magnified
Accuracy specification applies over
the
full
10
divisions.
Within ±2%
Within
±3%
When
in
X10
MAG, exclude first and last
50
ns
of
the
sweep
on
2-ns, 5-ns, 1 0-ns,
Within ±3%
Within
±4%
and 20-ns sweep rates.
±5%
over any two-division (or less)
portion
of
the
full 10 divisions. When
in X10 MAG, exclude first and last
magnified divisions when checking
2-ns, 5-ns, and 1 0-ns per division rates.
~±4
divisions.
Continuously variable between
At
least 2.5: 1.
calibrated settings. Extends
slowest A Sweep rate
to
at
least
1.25 s per division.
465B/DM44
Operators
REV A
AUG
1979
I
I
I
I
I
I
I
I
I

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 47E5A and is the answer not in the manual?

Tektronix 47E5A Specifications

General IconGeneral
BrandTektronix
Model47E5A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals