Getting started
The following t
able shows the output window (maximum and minimum levels)
for sine waveform when you change the load impedance (L). Load impedance
will affect the output window.
L=50Ω L=HighZ
AFG3011 / 30
11C
Maximum level
Minimum lev
el
(Maximum
amplitude)
10V-10V(20
Vp-p)
20 V -20 V (40
Vp-p)
AFG3021B / 3021C / 3022B / 3022C / 3051C / 3052C
Maximum level
Minimum level
(Maximum
amplitude)
5 V -5 V (10 Vp-p) 10 V -10 V (20 Vp-p)
AFG3101 / 3101C / 3102 / 3102C / 3151C / 3152C
Maximum
level
Minimum level
(Maximum
amplit
ude)
10 V -10 V (10 Vp-p) 10 V -10 V (20 Vp-p)
AFG325
1 / 3251C / 3252 / 3252C
Maximum level
Minim
um level
(Maximum
amplitude)
5V-5V
(5 Vp-p)
10 V -1
0 V (10 Vp-p)
Overheat Protection (AFG3011 / 3 011C Only)
The
instrument internal temperature is monitored in the AFG3011 and AFG3011C.
A warning message will appear if the internal temperature reaches a threshold
level, and signal output will automatically turn off. If the warning message
appears, check for the following conditions:
The ambient temperature requirement is being met.
The required cooling clearance is being met.
T
he instrument fan is working properly.
24 AFG3000 and AFG3000C Series User Manual