Performance verification
Error codes
If Diagnostics detects a malfunction, it displays the character string “Fail” and
the error code. Table 17 describes the Error code and related modules
Category of the Diagnostics Error Code.
Table 17 shows the error code and related modules reporting a failure.
Table 17: Error codes
Error code Description Related module
— Calibration Error Codes—
1101
CH1 Internal offset calibration failure A1x/A3x/A4x/A51/A61
1102
CH2 Internal offset calibration failure A12/A32/A42/A51/A61
1103
CH1 Output offset calibration failure A1x/A3x/A4x/A51/A61
1104
CH2 Output offset calibration failure A12/A32/A42/A51/A61
1105
CH1 Output gain calibration failure
A1x/A3x/A4x/A51/A61
1106
CH2 Output gain calibration failure
A12/A32/A42/A51/A61
1201
CH1 x3dB attenuator calibration failure A1x/A3x/A4x/A51/A61
1202
CH2 x3dB attenuator calibration failure A12/A32/A42/A51/A61
1203
CH1 x6dB attenuator calibration failure A1x/A3x/A4x/A51/A61
1204
CH2 x6dB attenuator calibration failure A12/A32/A42/A51/A61
1205
CH1 x10dB attenuator calibratio n failure A1x/A3x/A4x/A51/A61
1206
CH2 x10dB attenuator calibration failure A12/A32/A42/A51/A61
1207
CH1 x20dB 1 attenuator calibration failure A1x/A3x/A4x/A51/A61
1208
CH2 x20dB 1 attenuator calibration failure A12/A32/A42/A51/A61
1209
CH1 x20dB 2 attenuator calibration failure A1x/A3x/A4x/A51/A61
1210
CH2 x20dB 2 attenuator calibration failure A12/A32/A42/A51/A61
1211
CH1 Filter calibration failure A1x/A3x/A4x/A51/A61
1212
CH2 Filter calibration failure A12/A32/A42/A51/A61
1213
CH1 x20dB 3 attenuator calibration failure
A43
1301
CH1 Sine Flatness calibration failure A1x/A3x/A4x/A51/A61
1302
CH2 Sine Flatness calibration failure A12/A32/A42/A51/A61
1401
CH1 ASIC TINT calibration failure A1x/A3x/A4x/A51/A61
1402
CH2 ASIC TINT calibration failure A12/A32/A42/A51/A61
1403
CH1 ASIC SGEN calibration failure A1x/A3x/A4x/A51/A61
1404
CH2 ASIC SGEN calibration failure A12/A32/A42/A51/A61
1405
CH1 ASIC clock duty calibration failure A1x/A3x/A4x/A51/A61
1406
CH2 ASIC clock duty calibration failure A12/A32/A42/A51/A61
— Diagnostics Error Codes—
2100
Calibration data not found
A72
2101
Calibration data checksum
A72
AFG3000C and AFG3000 Series Specifications and Performance Verification 25