Performance verification
AFG325x / 325xC Measurement Range
Sine
+ 4.0 dBm 5.00 MHz
Reference ± 0.30
Sine
+ 4.0 dBm 25.00 MHz
Reference ± 0.50
Sine
+ 4.0 dBm 50.00 MHz
Reference ± 0.50
Sine
+ 4.0 dBm 100.00 MHz
Reference ± 1.00
Sine
+ 4.0 dBm 150.00 MHz
Reference ± 1.00
Sine
+ 4.0 dBm 240.00 MHz
Reference ± 2.00
Harmonics Distortion test
This test verifies the harmonic distortion using a spectrum analyzer.
1. Connect the arbitrary function generator to the spectrum analyzer as shownin
the following figure.
Figure 11: Harmonic distortion tests
2. Push the Default front-p
anel button and OK menu button to recall the arbitrary
function generator default setup.
3. Set up the arbitrary function generator using the following steps:
Select menu Setting Operation
Function
Sine Sine (front)
Frequency
20.00 kHz: AFG3011 /
3011C / 302xB / 302xC /
305xC
1.000 MHz: AFG310x /
310xC / 3151C / 3152C /
325x / 325xC
Frequency/Period (front)
Amplitude Units Vpp
TopMenu > Amplitude/Level Menu >
-more- > Units > Vpp
Amplitude 1.00 Vpp
Amplitude/High (front)
Channel 1 Output On On (front)
AFG3000C and AFG3000 Series Specifications and Performance Verification 59