Performance verification
AFG325x /
325xC Spectrum analyzer Measurement
50.00 MHz 10 MHz
300 MHz
20 MHz
600 MHz
20 kHz
20 kHz
MHz
MHz
dBc
dBc
< -41 dBc
100.00 MHz 10 MHz
300 MHz
20 MHz
600 MHz
20 kHz
20 kHz
MHz
MHz
dBc
dBc
< -35 dBc
200.00 MHz 10 MHz
300 MHz
20 MHz
600 MHz
20 kHz
20 kHz
MHz
MHz
dBc
dBc
< -29 dBc
240.00 MHz 10 MHz
300 MHz
20 MHz
600 MHz
20 kHz
20 kHz
MHz
MHz
dBc
dBc
< -28 dBc
Rise-Fall time test
This test
verifies the pulse rise time of the arbitrary function generator.
1. Connect the arbitrary function generator to the oscilloscope as shown in the
followi
ng figure.
Figure 14: Ri se-Fall time tests
2. Set up the arbitrary function generator using the following steps:
Select menu Setting Operation
Function
Square Square (front)
Frequency
2.00 MHz: AFG3011 /
3011C
10.00 MHz: AFG302xB
/ 302xC / 305xC / 310x /
310xC / 3151C / 3152C
/ 325x / 325xC
Frequency/Period (front)
Amplitude Units Vpp
TopMenu > Amplitude/Level Menu >
-more- > U nits > Vpp
Amplitude 1.0 Vpp
Amplitude/High (front)
AFG3000C and AFG3000 Series Specifications and Performance Verification 67