Clear and Sanitize Procedures
Memory Device
s
The following tables list the volatile and nonvolatile memory devices in the
standard instrument and listed options.
Table 1: Vol
atile Memory Devices for the DPO2000/B and MSO2000/B Series
Type and minimum size Function User modifiable
Data input
method Location To clear or sanitize
CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84
DDR memory: contains
the analog channel
waveform acquisition
data during operation
Only by the
processor system
N/A
Acquisition board
Remove power from the
instrument for at least
20 seconds
CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84
DDR memory: contains
the digital channel
waveform acquisition
data during operation
Only by the
processor system
N/A
Acquisition board
(MSO2000/B Series
only)
Remove power from the
instrument for at least
20 seconds
CMOS, SDRAM, 64MX16,
1G DDR2, 1.8V, 333MHZ;
MT47H64M16BT-3, FBGA84
DDR memory: contains
the instrument code and
data during operation
Only by the
processor system
N/A
Acquisition board
Remove power from the
instrument for at least
20 seconds
CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84
DDR memory: contains
the instrument code and
data during operation
Only by the
processor system
N/A
Acquisition board
(4-channel products
only)
Remove power from the
instrument for at least
20 seconds
DPO2000/B and MSO2000/B Series Declassification a nd Security Instructions 1