Commands Listed in Alphabetical Order
Examples
REM "This is a co
mment"
is a comment string that the oscilloscope will
ignore.
*RST (No Query Form)
Resets the oscilloscope to the factory default settings. The *RST command does
not alter the following:
Calibration data that affect device specications
The Output Queue
The Service Request Enable R egister setting
The Power-on status clea
r ag setting
Alias denitions
Stored settings
The *PUD? Response
Any o f the values associated with the DATA command.
Oscilloscope password
Group
Status and Error
Syntax
*RST
Related Commands
FACtory, RECAll:SETUp, SAVe:SETUp, *PSC, *RCL, *SAV
Arguments
None
Examples
*RST resets the oscilloscope settings to factory defaults.
*SAV (No Query Form)
Stores the state of the oscilloscope to a specied memory location. You can use
the *RCL command to restore the oscilloscope to this saved state at a later time.
Group
Save and Recall
Syntax
*SAV <N R1>
2-234 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Other manuals for Tektronix DPO2012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2012 and is the answer not in the manual?

Tektronix DPO2012 Specifications

General IconGeneral
BrandTektronix
ModelDPO2012
CategoryTest Equipment
LanguageEnglish

Related product manuals