Commands Listed in Alphabetical Order
Arguments
LESSThan sets t
he LIN data qualifier to less than.
MOREThan sets the LIN data qualifier to greater than.
EQUal sets the LIN data qualifier to equal.
UNEQual setstheLINdataqualifier to not equal.
LESSEQual sets the LIN data qualifier to less than or equal.
MOREEQual sets the LIN data qualifier to greater than or equal.
INrange sets the LIN data qualifier to in range.
OUTrange sets the LIN data qualifier to out of range.
Examples
TRIGGER:A:BUS:B1:LIN:DATA :QUALIFIER OUTRANGE sets the data qualifier
to out of range.
TRIGGER:A:BUS:B1:LIN:DATA :QUALIFIER? might return
TRIGGER:A:BUS:B1:LIN:DATA :QUALIFIER EQUAL indicating that the data
qualifier is set to equal.
TRIGger:A:BUS:B<x>:LIN:DATa:SIZe
Sets or returns the length of the data string in bytes to be used for LIN trigger.
Conditions
This command requires a DPO2AUTO application m odule.
Group
Bus
Syntax
TRIGger:A:BUS:B<x>:LIN:DA Ta:SIZe <NR1>
TRIGger:A:BUS:B<x>:LIN:DA Ta:SIZe?
Arguments
<NR1> is the size of the data string in bytes.
Examples
TRIGGER:A:BUS:B1:LIN:DATA :SIZE 8 sets the data size to 8 bytes.
TRIGGER:A:BUS:B1:LIN:DATA :SIZE? might return
TRIGGER:A:BUS:B1:LIN:DATA :SIZE 1 indicating that the data size is 1 byte.
TRIGger:A:BUS:B<x>:LIN:DATa:VALue
Sets or returns the binary data string to be used for LIN trigger condition if trigger
condition is ID or IDANDDATA.
2-306 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual