Commands Listed in Alphabetical Order
WFMOutpre:XI
Ncr? (Query O nly)
Returns the horizontal point spacing in units of WFMOutpre:XUNit for the
waveform speci ed by the DATa:SOUrce command. This value corresponds to
the sampling
interval. The query command will time out and an error will be
generated if the waveform specied by DATa:SOUrce is not turned on.
Group
Waveform Transfer
Syntax
WFMOutpre:XINcr?
Related Commands
DATa:SOUrce, WFMOutpre:XUNit?
Examples
WFMOUTPRE:XINCR? might return :WFMOUTP RE:XINCR 10.0000E-6
indicating that the horizontal sampling interval is 10 µs/point.
WFMOutpre:XUNit? (Query Only)
Returns the horizontal units for the waveform specied by the DATa:SOUrce
command. The query command will time out and an e rror will be generated if the
waveform specied by DATa:SOUrce is not turned on.
Group
Waveform Transfer
Syntax
WFMOutpre:XUNit?
Related Commands
DATa:SOUrce
Examples
WFMOUTPRE:XUNIT? might return :WFMOUTPRE:XUNIT "HZ" indicating that
the horizontal units for the waveform are in Hertz.
WFMOutpre:XZEro? (Query Only)
Returns the time coordinate of the rst point in the outgoing w aveform.
This value is in units of WFMOutpre:XUNit?. The query command will t
ime
out and an error will be generated if the waveform specied by DATa:SOUrce
is not turned on.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-371

Other manuals for Tektronix DPO2012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2012 and is the answer not in the manual?

Tektronix DPO2012 Specifications

General IconGeneral
BrandTektronix
ModelDPO2012
CategoryTest Equipment
LanguageEnglish

Related product manuals