Trigger Setup
Untriggered display Triggered display
Trigger Modes
The trigger mode determines how the instrument behaves in the absence of a trigger event:
Normal trigger mode enables the instrument to acquire a waveform only w hen it is triggered. If no trigger occurs, the last
waveform record acq uired remains on the display. If no last waveform exists, no w aveform is displayed.
Auto trigger mode en ables the instrument to acquire a waveform even if a trigger does not occur. Auto mode uses a timer that
starts when the acquisition is started , and the pretrigger information is obtained. If a trigg er event is not detected before the
timer times out, the instrument forces a trigger. The length of time it waits for a trigger event depends on the time base setting.
Auto mode, when forcing triggers in the absence of valid triggering events, does not synchronize the waveform on the display. The
waveform will a ppear to roll across the screen. If valid triggers occur, the display will become stable.
MSO4000 and DPO4000 Series Oscilloscopes User Manual 131

Table of Contents

Other manuals for Tektronix DPO4032

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO4032 and is the answer not in the manual?

Tektronix DPO4032 Specifications

General IconGeneral
BrandTektronix
ModelDPO4032
CategoryTest Equipment
LanguageEnglish

Related product manuals