Adjustment Procedures
This chapter contains information about instrument adjustment.
Adjustment Interval
The voltage and timing refe rences inside the instrument are very stable over time
and should not need routine adjustment.
If the instrument fails the performance tests in the Specifications and Performance
Ver i fication manual, adjustment may be required. This manual was provided with
your product, and can also be found at www.tektronix.com/manuals by searching
foryourproduct.
If periodic calibration is one of your requirements, a general rule is to verify
performance and make adjustments (only if needed) every 2000 hours of operation
or once a
year if the instrument is used infrequently.
Adjustment After Repair
After r
emoval and replacement of a module, you must perform the Performance
Ver i fication procedure, found in the Specifications and Performance Verification
manual. This manual was provided with your product, and can also be found at
www.tektronix.com/manuals by searching for your product.
If the instrument fails the Performance Verification tests, it must be returned to
Tektronix for calibration.
Adjustment
Adjustment can be performed only by a Tektronix Service Center. See Contacting
Tektronix, following the title page in this manual, for information on contacting
Tektronix Service Support.
MSO70000/C, DSA70000B/C, DPO70000B/C, D PO7000, MSO5000, DPO5000 Series 3–1