Page #96 background imageLoading...
Page #96 background image
Pinpoint trigge
rs
Checking trigger status
Check the trigger status from the status lights on the front panel or from the readout.
Check the ARM, READY, and TRIG'D
front-panel controls to determine the trigger
status.
If TRIG'D is on, the i nstrument has
recognized a valid trigger and is lling
the posttrigger part of the waveform.
If READY is on, the instrument can
accept, and is waiting for, a valid trigger
to occur. Pretrigger data has been
acquired.
If ARM is on, the trigger circuitry is
lling the pretrigger part of the waveform
record.
If both TRIG'D and READY are on,
a valid A event trigger has been
recognized and the instrument is waiting
for a delayed trigger. When a delayed
trigger is recognized, the posttrigger part
of the delayed waveform will ll.
If ARM, TRIG'D, and READY are off,
acquisitions have stopped.
MSO/DPO70
000DX, MSO/DPO70000C, DPO7000C Serie s
To quickly determine the settings of some
key trigger parameters, check the Trigger
readout at the bottom of the display. The
readouts differ for edge and the advanced
triggers:
1. A trigger source = Ch1
2. Trigger slope = rising edge
3. Trigger level
4. Time base
5. Check the ARM, READY, and TRIG'D
readout to determine the trigger status.
6. Check the Auto/Normal readout to
determine the trigger mode.
74 MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO 5000B Series U ser M anual

Table of Contents

Other manuals for Tektronix DPO7054C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO7054C and is the answer not in the manual?

Tektronix DPO7054C Specifications

General IconGeneral
BrandTektronix
ModelDPO7054C
CategoryTest Equipment
LanguageEnglish

Related product manuals