REMOTE CONTROL
43
Parameter <voltage>
<NRf> 0% ~ 105% of the rated output
voltage.
<current>
<NRf> 0% ~ 105% of the rated output
current.
MIN
0 volts/0 amps
MAX
Maxium value for the present range.
Return parameter
<NRf> Returns the voltage and current.
Example APPL 5.05,1.1
Sets the voltage and current to 5.05V and 1.1A.
Query Example APPL?
+5.050, +1.100
Returns voltage (5.05V) and current (1.1A) setting.
Display Commands
DISPlay:MENU[:NAME] ................................................... 43
DISPlay[:WINDow]:TEXT:CLEar .................................... 44
DISPlay[:WINDow]:TEXT[:DATA] ................................. 44
DISPlay:BLINk ..................................................................... 45
DISPlay:MENU[:NAME]
Description
The DISPlay MENU command selects a screen
menu or queries the current screen menu.
Syntax
Query Sytax
DISPlay:MENU[:NAME] <NR1>
DISPlay:MENU[:NAME]?
Parameter/
Return parameter
<NR1>
Description
0
Measurement-Voltage / Measurement-
Current
1
Measurement-Voltage / Measurement-Power
2
Measurement-Power / Measurement-Current
3
Set Menu
4
OVP / OCP Menu
5~99
Not Used.