REMOTE CONTROL
57
[SOURce:]CURRent:PROTection:STATe
Description
Turns OCP (over-current protection) on or off.
Syntax [SOURce:]CURRent:PROTection:STATe {0|1|OFF|ON}
Query Syntax [SOURce:]CURRent:PROTection:STATe?
Parameter/Return
0
<NR1> Turns the buzzer off.
OFF
Turns the buzzer off.
1
<NR1> Turns the buzzer on.
ON
Turns the buzzer on.
Return parameter
<Bool>
Returns bleeder resistor status (0 or 1).
Example SOUR:CURR:PROT:STAT OFF
Turns OCP off.
[SOURce:]CURRent:SLEW:RISing
Description
Sets or queries the rising current slew rate. This is
only applicable for CC slew rate priority mode.
Syntax
Query Syntax
[SOURce:]CURRent:SLEW:RISing {<NRf>|MIN|MAX}
[SOURce:]CURRent:SLEW:RISing? [MIN|MAX]
Parameter/Return
<NRf>
0.01A/s~200% (2260B-30-36)
0.1A/s~200% (2260B-30-72)
0.01A/s~200% (2260B-80-13)
0.01A/s~200% (2260B-80-27)
MIN
Minimum rising current slew rate.
MAX
Maximum rising current slew rate.
Example SOUR:CURR:SLEW:RIS 72
Sets the rising current slew rate to 72A/s.
[SOURce:]CURRent:SLEW:FALLing
Description
Sets or queries the falling current slew rate. This is
only applicable for CC slew rate priority mode.
Syntax [SOURce:]CURRent:SLEW:FALLing {<NRf>|MIN|MAX}
Set
Query
Set
Query
Set
Query