REMOTE CONTROL
75
*CLS
Description
The *CLS command clears the Standard Event
Status, Operation Status and Questionable Status
registers. The corresponding Enable registers in
each of the above registers are not cleared.
If a <NL> newline code immediately precedes a
*CLS command, the Error Que and the MAV bit in
the Status Byte Register is also cleared.
Syntax *CLS
*ESE
Description
Sets or queries the Standard Event Status Enable
register.
Syntax
Query Syntax
*ESE <NR1>
*ESE?
Parameter <NR1>
0~255
Return parameter
<NR1>
Returns the bit sum of the Standard Event
Status Enable register.
*ESR
Description
Queries the Standard Event Status (Event) register.
The Event Status register is cleared after it is read.
Query Syntax *ESR?
Return parameter
<NR1>
Returns the bit sum of the Standard Event
Status (Event) register and clears the register.
*IDN
Description
Queries the manufacturer, model name, serial
number, and firmware version of the unit.
Query Syntax *IDN?
Set
Set
Query
Query
Query
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Tektronix Keithley 2260B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2260B Series and is the answer not in the manual?

Tektronix Keithley 2260B Series Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2260B Series
CategoryPower Supply
LanguageEnglish

Related product manuals