The *CLS command clears the Standard Event
Status, Operation Status and Questionable Status
registers. The corresponding Enable registers in
each of the above registers are not cleared.
If a <NL> newline code immediately precedes a
*CLS command, the Error Que and the MAV bit in
the Status Byte Register is also cleared.
Syntax *CLS
*ESE
Description
Sets or queries the Standard Event Status Enable
register.
Syntax
Query Syntax
*ESE <NR1>
*ESE?
Parameter <NR1>
0~255
Return parameter
<NR1>
Returns the bit sum of the Standard Event
Status Enable register.
*ESR
Description
Queries the Standard Event Status (Event) register.
The Event Status register is cleared after it is read.
Query Syntax *ESR?
Return parameter
<NR1>
Returns the bit sum of the Standard Event
Status (Event) register and clears the register.
Description
Queries the manufacturer, model name, serial
number, and firmware version of the unit.
Query Syntax *IDN?