Why can't the DMM7510 read my USB flash drive?
Verify that the flash drive is formatted with the FAT32 file system. The DMM7510 only supports FAT
and FAT32 drives using Master Boot Record (MBR).
In Microsoft
®
Windows
®
, you can check the file system by checking the properties of the USB
flash drive.
Higher capacity USB drives take longer to be read and loaded by the instrument.
How do I change the command set?
You can change the command set that you use with the DMM7510. The remote command sets that
are available include:
• SCPI: An instrument-specific language built on the SCPI standard.
• TSP: A scripting programming language that contains instrument-specific control commands that
can be executed from a stand-alone instrument. You can use TSP to send individual commands
or use it to combine commands into scripts.
If you change the command set, reboot the instrument.
You cannot combine the command sets.
As delivered from Keithley Instruments, the DMM7510 is set to work with the SCPI command set.
To set the command set from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. Select the appropriate Command Set.
You are prompted to confirm the change to the command set and reboot the instrument.