Section
7: Sampling temperature at a set time interval Model DMM7510 7½ Digit Multimeter
User's Manual
7-4 DMM7510-900-01 Rev. C / September 2019
Send the commands in the following table to set up and run this example application.
Command Description
*RST
Reset the DMM7510.
MeasCount = 1440
MeasInterval = 60
Define variables to make a measurement
every 60 seconds for 24 hours (1440 times).
:SENSe:FUNCtion "TEMPerature"
:SENSe:TEMPerature:TRANsducer FRTD
:SENSe:TEMPerature:RTD:FOUR PT100
:SENSe:TEMPerature:ODETector ON
:SENSe:TEMPerature:NPLCycles 1
:SENSe:TEMPerature:OCOMpensated ON
:DISPlay:SCReen GRAPh
:TRIGger:TIMer1:DELay MeasInterval
:TRIGger:TIMer1:STARt:STIMulus NOTify1
:TRIGger:TIMer1:STARt:GENerate OFF
:TRIGger:TIMer1:STATe ON
:TRIGger:BLOCk:WAIT 1, DISPlay
:TRIGger:BLOCk:NOTify 2, 1
:TRIGger:BLOCk:MDIGitize 3, "defbuffer1"
:TRIGger:BLOCk:WAIT 4, TIMer1
:TRIGger:B
LOCk:BRANch:COUNter 5, MeasCount,
2
:INIT
Set the instrument to measure temperature.
Set the transducer type to 4-wire RTD.
Set the RTD to type PT100.
Enable the detection of open leads.
Set NPLC to 1.
Enable offset compensation.
Switch to the GRAPH swipe screen.
Set up timer 1 with delay time of
MeasInterval.
Set the notify 1 event to trigger the start of
the timer.
Generate a timer event once when the timer
delay elapses.
Enable the timer.
Set trigger model block 1 to wait for
front-panel TRIGGER key press.
Set block 2 to send NOTIFY1 to start
the timer.
Set block 3 to make a measurement and
store it in defbuffer1.
Set block 4 to wait for the timer 1 event after
the delay elapses.
Set block 5 to go to block 2
MeasCount times.
Initiate the trigger model.
Using TSP
The following TSP code is designed to be run from Keithley Instruments Test Script Builder (TSB).
TSB is a software tool that is available from tek.com/keithley
. You can install and use TSB to write
code and develop scripts for TSP-enabled instruments. Information about how to use TSB is in the
online help for TSB and in the “Introduction to TSP operation” section of the Model DMM7510
Reference Manual.
To use other programming environments, you may need to make changes to the example TSP code.
By default, the DMM7510 uses the SCPI command set. You must select the TSP command set
before sending TSP commands to the instrument.

Table of Contents

Other manuals for Tektronix Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley DMM7510 and is the answer not in the manual?

Tektronix Keithley DMM7510 Specifications

General IconGeneral
BrandTektronix
ModelKeithley DMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals