Page #11 background image
Memory devices
The following sections list the volatile and nonvolatile memory devices in the
standard instrument and listed options.
Volatile memory devices
DDR3
Type and size 1 GB standard
Function Holds active analog acquisition data
Type of user information
stored
User data
Backed up by battery? No
Method of modification Indirectly
Data input method Firmware operations
Location Main board
User accessible No
To clear Remove power from the instrument for at least 20 seconds.
Process to sanitize Remove power from the instrument for at least 20 seconds.
SDRAM
Type and size SDRAM 128 GB (standard)
Function Holds active digital acquisition data
Type of user information
stored
User data
Backed up by battery? No
Method of modification Indirectly
Data input method Firmware operations
Location Main board
User accessible No
To clear Remove power from the instrument for at least 20 seconds.
Process to sanitize Remove power from the instrument for at least 20 seconds.
SDRAM
Type and size SDRAM 32 MB (standard)
Function ASIC execution memory
Type of user information
stored
User data and settings
Backed up by battery? No
Method of modification indirectly
Data input method Firmware operations
MDO3000 Series Declassification and Security Instructions 1

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals