Acquisition
Using Roll Mode
Roll mode gives a display similar to a strip chart recorder for low-frequency signals. Roll mode displays acquired data points
without waiting for the acquisition of a complete waveform record.
1. Select Horiz/Acq >
Horizontal/Acquisition Setup....
2. If not selected, click the Acquisition tab.
Click Auto to turn on Roll mode.
NOTE. Roll
mode requires Sample, Peak
Detect, or H i Res acquisition mode.
3. To stop acquisitions in Roll mode:
If you are not in Single Sequence,
push RUN/STOP to stop Roll mode.
If you are in Single Sequence, Roll
mode acquisitions stop automatically
when a complete record is acquired.
MSO/DPO70000DX, MSO/DPO70000C, and DPO7000C Series
MSO5000B and DPO5000B Series
Quick tips
Switching to Envelope, Average, or WfmDB acquisition mode will turn off Roll mode.
Roll mode is disabled when you set the horizontal scale to 50 ms per division or faster.
50 MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO 5000B Series U ser M anual

Table of Contents

Other manuals for Tektronix MSO5104B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO5104B and is the answer not in the manual?

Tektronix MSO5104B Specifications

General IconGeneral
BrandTektronix
ModelMSO5104B
CategoryTest Equipment
LanguageEnglish

Related product manuals