Command Descriptions
2–66
TDS Family Oscilloscope Programmer Manual
POSition
CH
<x>
:
?
<Space> <NR3>
<NR3> is the desired position, in divisions from the center graticule. The range is
±5 divisions.
CH2:POSITION 1.3E+00
positions the channel 2 input signal 1.3 divisions above the center of the display.
CH1:POSITION?
might return -1.3E+00, indicating that the current position of channel 1 is at
–1.3 divisions.
CH<x>:PRObe? (Query Only)
Returns the attenuation factor of the probe that is attached to the specified
channel.
Vertical
CH<x>:PRObe?
CH
<x> :
PRObe
?
<NR3>
CH4:PROBE?
might return 100.0E-3 for a 10x probe.
CH<x>:SCAle
Sets or queries the vertical gain of the specified channel. This is equivalent to
setting Fine Scale in the Vertical menu or adjusting the front-panel Vertical
SCALE knob.
Vertical
CH1:VOLts
Arguments
Examples
Group
Syntax
Returns
Examples
Group
Related Commands

Table of Contents

Other manuals for Tektronix TDS 700A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700A and is the answer not in the manual?

Tektronix TDS 700A Specifications

General IconGeneral
BrandTektronix
ModelTDS 700A
CategoryTest Equipment
LanguageEnglish

Related product manuals