TDS1000 and TDS2000 S eries Digital Storage Oscilloscopes Service Manual
1-1
Specifications
These specifications apply to all TDS1000- and TDS2000-series oscilloscopes.
To verify that an oscilloscope meets specifications, it must first meet the
following conditions:
H The oscilloscope must have been operating continuously for twenty minutes
within the specified operating temperature.
H You must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5 °C.
H The oscilloscope must be within the factory calibration interval of one year.
Specifications begin in Table 1--1. All specifications are guaranteed unless noted
“typical.” Specifications that are marked with the n symbol are checked in the
chapter Performance Verification.
Table 1- 1: Oscilloscope specifications
Acquisition
Acquisition Modes Sample, Peak Detect, and Average
Acquisition Rate,
typical
Up to 180 waveforms per second, per channel (Sample acquisition mode, no measurements)
Single Sequence Acquisition Mode Acquisition Stops After
Sample, Peak Detect Single acquisition, all channels simultaneously
Average N acquisitions, all channels simultaneously. N is selectable from 4,
16, 64, and 128
Inputs
Input Coupling DC, AC, or GND
Input Impedance, DC
Coupled
1MΩ ±2% in parallel with 20 pF ±3pF
P2200 Probe
Attenuation
1X, 10X
Supported Probe
Attenuation Factors
1X, 10X, 100X, 1000X